Search in catalog

Česky English Deutsch Français Russian Polski Español
Breadcrumb navigation

Home > News > Mash Bath with automatic distilled water refilling

Mash Bath with automatic distilled water refilling



Due date:
20.11.2008

 

Mashing Equipment is designed for laboratory use in breweries and malthouses. It aids in the determination of malt extract. The equipment allows one to utilize the CONGRESS, HARTONG 20, HARTONG 45, HARTONG 65, HARTONG 80, ASBC, PROFILE or THERMOSTAT methodology of testing. The Thermostat method maintains a constant temperature for a specific time. The Profile method allows the user to program the variables to suite their needs.

Two versions are available:

Standard version: all the above mentioned methods run fully automatically as an alarm lets the operator know the necessity for example to sample for saccharification, to refill the distilled water (which is preheated in test tubes directly in the Mash Bath) into the beakers with the mash at predetermined intervals.

Automatic version: with automatic distilled water refilling: all the above mentined methods run fully automatically including automatic preheated distilled water refilling.

The mashing equipment is equipped with a microcomputer, a thermometer Pt 100, an electromagnetic valve for the automatic cooling-down of mash bath, an LCD double line display, a membrane keyboard and controlling software. The mashing unit can be equipped with a serial boundary at the customers request. This serial boundary makes communication possible with a PC. The Mash Bath equipped with a serial boundary is indicated as monitoring. This is an advantage for customers that are on an ISO 9001 compliant system.

Models:

  • Standard R4- four beakers
  •  Standard R4 monitoring - four beakers, equipped with a interface RS232  
  • Standard R8 - eight beakers
  • Standard R8 monitoring - eight beakers, equipped with a interface RS232
  • Automatic R8 monitoring - eight beakers, with automatic distilled water refilling, equipped with a interface  RS232
  • Standard R12 - twelve beakers
  • Standard R12 monitoring - twelve beakers, equipped with a interface RS232
  • Automatic R12 monitoring-  twelve beakers, with automatic distilled water refilling, equipped with a interface RS232

 

Data represented on LCD display:
  • type of method: Hartong 20, Hartong 45, Hartong 65, Hartong 80, Congress, ASBC, Thermostat, Profile
  • instantaneous temperature of bath
  • time since the begining of test

 

Adjustable data:
  • choice of methods: Hartong 20, Hartong 45, Hartong 65, Hartong 80, Congress, ASBC, Thermostat, Profile
  • temperature range: from 200C to 950C
  • mixer speed: 0, 100, 200 R.P.M.
  • time
  • accuracy of temperature adjustment: +-0,010C
  • accuracy of regulation: to 0,2C
  • accuracy of time adjustment: 1 sec

 

Signalization:

  • acoustic and optical alarm

 

Electric data:

  • voltage 230V/50 Hz
  • Note: by request 110V/60Hz

Mashing Bath.pdf